Time-electronics 1048 Voltage-Current-Loop Calibrator Manual do Utilizador Página 10

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 23
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 9
10
V1.2 30/12/10 Page | 10 1048 Technical Manual
3.7. Sink - transmitter simulator (TX SIM)
In the Sink function, the unit simulates a process field transmitter as part of a loop testing
and calibration operation. The unit takes its excitation power from the loop and acts as a
variable resistance to adjust the loop current to the value shown on the units display. The
excitation voltage must be in the range 3 to 50 volts dc.
The NORM/OFF/REV switch in not functional in the SINK mode.
Set function switch to TX SIM
Set OFF/V/mA switch to mA
Set range switch to the 22 position
Set output switch to NORM
Set multi-turn output control to minimum (anticlockwise)
Connect the unit in the loop
observing correct polarity. Set the
current in the loop by adjusting the
multi-turn control (coarse) and the
INC/DEC buttons (fine)
Vista de página 9
1 2 ... 5 6 7 8 9 10 11 12 13 14 15 ... 22 23

Comentários a estes Manuais

Sem comentários